Honeywell Sensotec Frequently Asked Questions

What are the advantages and disadvantages of having an internal amplifier on a pressure detector?


Systems without internal amplification must receive supply voltage from an external source, and must send low level signals (ie. millivolt signals) back to the amplifying source, sometimes from a long distance.

External Amp
External Amp sends supply power a long distance and receives the small (30 mV) transducer signal from a long distance

The long distances experienced in some applications can create problems for unamplified transducers. The output signal can be degraded by external noise sources.

Internal amplification
Transducer signals may become distorted Internal amplification can minimize these effects.

Internal amplifiers are housed in the same unit as the sensor.The transducer output is amplified and therefore a much larger signal 0 - 10V.

Internally amplified transducer
Internally amplified transducers provide a high level output

This ensures that the induced electrical noise is a couple of orders of magnitude smaller than the amplified signal and therefore creates a higher signal to noise ratio. The larger output also allows A/D converters to create a higher resolution input. The amplifier outputs are low impedance, and internal amplifiers don't contribute noticeably to system inaccuracy.

Amplified output
Amplified output with a much higher signal to noise ratio

Internal amplifiers may not be feasible under certain conditions. Specifically, the circuitry in the amp can not be subjected to extreme temperatures. If sensor placed in a location inaccessible to users (hazardous environment, small space, long distance), zero and span adjustments may not be able to be tweaked when needed. Internal amplifiers increase the overall size of the unit, which may be concern in some applications.



Offline
pdf document Download our new FAQ document for a hardcopy of more than 100 FAQs at your fingertips.

Online
Select a category below, or use the keywords section to explore our online "experience" database.
Categories:
  1. Accelerometers
  2. Calibration
  3. Calibration Class Load Cells
  4. Load Cells
  5. Measurement General
  6. Pressure Sensors
  7. Torque
Keyword Search:   
                                   Please click button to search.

Certificates | Privacy Statement | Search

1-800-867-3888   |  Request Information

Copyright 2004-2008 Honeywell International Inc. All rights reserved.