The special requirements of CIP procedures can introduce
inaccuracies in temperature and pressure readings.
Clean-in-place (CIP) procedures require in situ cleaning
of all the internal components of a process system, including
the sensor head, after each batch run to prevent bacterial
growth and contamination so as to maintain the highest standard
of product quality. This practice is common to a variety
of industries, such as food and beverage; pharmaceuticals;
fats and oils; paint and ink; cosmetics and perfumes; prepared
animal feed;chemical manufacturing; and clean gas systems.
CIP units must survive high temperatures and cope with
rapid temperature changes that can impair sensor accuracy.
This specification is generally defined as the sum of the
sensor's linearity, hysteresis, and repeatability.
A unit's accuracy is affected by temperature effect, which
is seen as an off set of the zero and span of the sensor.
Both are usually expressed as a +/- percentage change in
the rated output per °F change in ambient temperature
over the compensated temperature range of the particular
model. This could, for example, be expressed as 0.0025/
°F change on both zero and span.
The process engineer uses these temperature effect numbers
to calculate the applied accuracy of a given sensor, taking
into account the actual temperature of the application.
TEMPERATURE
Temperature-induced expansion is a problem for all sensors
and instruments. Sensors are typically constructed of various
types of metal welded together to form a unit. To achieve
optimal thermal stability, these metals must be matched
in terms of their thermal expansion coefficients, and high-tech
welding processes must be used to join them. When extra
large metal components (e.g., a sanitary pressure flange)
are part of the sensing unit, the thermal expansion problem
becomes considerably more complicated.
The temperature gradient that develops across a transducer
body defies the compensating techniques used by most sensor
manufacturers. The reason is that the temperature compensation
is electrical and located electrical and located at the
sensor's back end with the rest of the electronic components.
Therefore, while the front end of a sensor may see temperatures
of 200°F-400°F, the electronic compensation components are
in a very different climate (see Figure I).
 |
Figure 1.
As the process medium comes into contact with the
sensor diaphragm, the sensor experiences a temperature
gradient that heats first the coupling assembly and
then the sensor to different levels at different locations. |
This temperature gradient effect also influences response
to temperature change, a particularly thorny design problem.
If the nature of the application permits the unit to stabilize
at some temperature before a reading is taken, the transducer's
accuracy rating is reliable. In the real world, however,
most transducers take from 1 to 2½ hr. to stabilize
at a new temperature when they move from ambient to above
200 °F, and few processes are slow enough to allow that
amount of time for stabilization.
Sensor manufacturers have attacked this problem in two
ways. Smart transmitters were developed with on-board electronic
temperature compensation, which is one of the factors that
make them much more expensive than the conventional variety.
A fairly recent innovation combines basic sensor technology
and a clever mechanical design (see Figure 2) that permits
non-smart sensors to track temperature changes as fast as
200°/ min., which is more than adequate for many processes.
These devices are useful when it is necessary to precisely
control some time-related part of the process where the
temperatures are changing, e.g., adding ingredients at the
proper time and/or in the proper quantity, or opening and
closing valves or doors.
|
Figure 2.
An ULTRA Series sensor with a 1 ½-in. clean-in-place
(C1P) flange was electronically given an intentional
zero offset to simulate a constant pressure. As the
temperature of the process medium was increased, the
average temperature of the sensor rose, causing a
shift in the simulated pressure reading. The observed
change, called zero offset, illustrates temperature-induced
sensor inaccuracies. Approximately 50 min. into this
test, the sensor was removed from the ClP clamp and
immediately immersed in room temperature water. A
sharp decline in both the process medium and the sensor
temperature readings, and a concurrent decline in
the zero offset line, show the sensor's almost instantaneous
response to the extremely rapid temperature change.
Most sensors would exhibit the same response, but
over a 15-30 min. time period instead of only a few
minutes. |
PRESSURE
The term sanitary (hygienic in Europe) denotes a portion
of the CIP market that makes products for personal use or
ingestion by human beings. The Sanitary Standards Council,
in conjunction with the FDA, has developed standards that
must be met for a manufacturer to display the 3A label.
These standards pertain to both the mechanical design and
the fluids that transmit pressure inside the sensor.
Diaphragms (see Photo 1), for example, must fit flush and
have no crevices that could capture and retain impurities.
| |
Photo 1. The flush-mounted, smooth-surfaced
diaphragm of this 3A-rated pressure sensor is sensitive
enough to provide a timely response to changes in
temperature and pressure with 0.15% F.S. accuracy.
Although massive, the diaphragm is sensitive enough
to measure pressure from 600 psi down to 3.5 psi.
The sensor assembly withstands the 400E°F temperatures
associated with ClP procedures.
|
Temperature compensation is also complicated by the 3A
guideline dictating the use of an edible fluid to transmit
pressure sensed on the flush diaphragm of the sensing element.
This fluid is somewhat compressible, and has a thermal expansion
coefficient quite different from that of any of the metals
used in sensor construction. Only within the past year has
it become possible to substantially minimize the impact
of the edible oil.
Furthermore, if the sensor manufacturer does not provide
the proper controls, these fluids can fill up with air molecules
that separate from the fluid at process temperatures. As
bubbles are formed, the gas expands or is compressed as
space allows, resulting in erroneous pressure readings.
The edible pressure transmission oil introduces yet another
variable, called the orientation effect. Each oil has a
given density (weight), which the sensor will detect as
a pressure. The amount of pressure exerted by the oil depends
on how the sensor is installed in the process piping. If
a unit is installed with the diaphragm up, for example,
the total weight of the oil is on the sensor; if the unit
is positioned with the sensing head down, it sees no oil
weight. Different degrees of angle in a horizontal mounting
will produce varying weights. The user will see this phenomenon
as a zero shift.
So, as vibration or system maintenance changes the sensor’s
orientation, a skilled technician or an engineer will have
to go to the sensor site (which is often remote and in hard-to-access
spaces) and adjust a potentiometer on the unit to eliminate
the zero shift. The degree of orientation effect is determined
by sensor design, and it varies from one manufacturer to
another. Recent advances have made this effect almost negligible,
with the availability of units with 0.03 psi/g maximum error
due to orientation. There are few processes that would be
bothered by this small variation. |